Table 3. Performance Test Points:
Current Range 0 A to 10 A (100 mV/A)
DC Measurement:
5520A
Settings
DC Amps
Measured
Low Limit
Output
High Limit
Output
0.1 A 1 A 92 mV 108 mV
0.5 A 5 A 480 mV 520 mV
0.9 A 9 A 868 mV 932 mV
AC Measurement:
5520A
Settings
RMS Amps
Measured
Low Limit
Output
High Limit
Output
0.1 A, 60 Hz 1 A 92 mV 108 mV
0.3 A, 400 Hz 3 A 286 mV 314 mV
0.5 A, 2 kHz 5 A 465 mV 535 mV
0.6 A, 4 kHz 6 A 559 mV 641 mV
Table 4. Performance Test Points:
Current Range 0 A to 100 A (10 mV/A)
DC Measurement:
5520A
Settings
DC Amps
Measured
Low Limit
Output
High Limit
Output
1 A 10 A 95.5 mV 104.5 mV
3 A 30 A 287.5 mV 312.5 mV
5 A 50 A 439.5 mV 560.5 mV
7 A 70 A 615.5 mV 784.5 mV
9 A 90 A 765.0 mV 1035.0 mV
AC Measurement:
5520A
Settings
RMS Amps
Measured
Low Limit
Output
High Limit
Output
1 A, 5 kHz 10 A 92.5 mV 107.5 mV
2 A, 1 kHz 20 A 185.5 mV 214.5 mV
3 A, 400 Hz 30 A 287.5 mV 312.5 mV
5 A, 400 Hz 50 A 439.5 mV 560.5 mV
7 A, 60 Hz 70 A 595.0 mV 805.0 mV