Extended Accuracy:
For other frequencies, refer to the appropriate input current range and add the error
listed below to the "Basic Accuracy" error.
Input Load Impedance
(of host instrument) .......................... >1 M in parallel with up to 100 pF
Useful Bandwidth (-3 dB) ................. 0 kHz to 100 kHz
Rise or Fall Time .............................. <4 μsec
Output noise level
Range 10 mV/A typical............. 480 μV pk-pk
Range 100 mV/A typical........... 3 mV pk-pk
Max. nondestructive current............. 0 kHz to 2 kHz 140 A peak
2 kHz to 10 kHz 110 A peak
10 kHz to 20 kHz 70 A peak
20 kHz to 50 kHz 30 A peak
50 kHz to 100 kHz 20 A peak
Temperature coefficient.................... 2000 ppm/°C max. for temperature
from 0 °C to 50 °C (32 °F to 132 °F)
Performance Verification
Verify Probe accuracy with the test setup shown in Figure 5. Required test
equipment is defined in Table 1. Toroid coil construction is illustrated in
Ta bl e 2 .
To verify the Probe accuracy:
1. Set up connection.
2. Make the checks called for in Table 3 (100 mV/A).
3. Make the checks called for in Table 4 (10 mV/A).
Frequency
Additional Error
Range 100 mV/A Range 10 mV/A
1 kHz to 5 kHz
5 kHz to 20 kHz
>20 kHz
3 %
12 %
not specified
3 %
12 %
not specified
Table 1. Required Test Equipment
Required Recommended
AD/DC Calibrator Fluke 5520A
Digital Multimeter (DMM) Fluke 45
Small insulated screwdriver Spectrol
Banana-to-BNC Adapter Fluke PM9081/001
10-turn Toroid Coil See Figure 6
80i-110s.book Page 9 Monday, December 5, 2016 12:12 PM