EasyManua.ls Logo

Fluke 825A - Troubleshooting

Fluke 825A
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
825А
(5)
After
chopper
is
thoroughly
dry,
replace
chopper
cover
and
install
in
instrument.
NOTE
If
the
chopper
drive
unit
is
unplugged
while
the
instrument
is
on,
the
chopper
drive
unit
will
not
operate
when
reassembled.
How-
ever,
the
chopper
drive
unit
will
operate
if
the
instrument
is
turned
off
and
then
turned
back
on.
(6)
Adjust
УТУМ
ZERO
ADJ
control
R232
for
zero
meter
deflection
after
a
20
minute
warmup
period.
4-3.
TROUBLESHOOTING
a.
GENERAL.
To
assist
in
localizing
most
troubles
which
might
occur,
a
troubleshooting
chart
(figure
4-2)
lists
the
causes
and
remedies
for
a
number
of
symp-
toms.
Since
failure
of
the
825A
is
usually
due
to
tube
failure,
it
is
recommended
that
tube
substitution
be
tried
before
any
other
tests
are
made.
However,
the
825A
is
conservatively
designed
and
should
require
very
infrequent
tube
replacement.
At
certain
times,
a
check
of
the
DC
pin
voltages
in
the
tubes
(figure
4-6)
is
also
useful.
As
an
aid
to
troubleshooting,
all
components
may
be
located
by
referring
to
the
figures
shown
in
Sec-
tion
V.
An
understanding
of
the
theory
of
operation
(Section
III)
and
frequent
reference
to
the
functional
schematic
will
be
of
great
value
in
troubleshooting.
SYMPTOM
PROBABLE
CAUSE
REMEDY
Drift
of
the
500
volt
DC
reference
supply
evi-
denced
by
the
continual
need
for
resetting
with
the
CALIBRATE
control.
Cannot
calibrate
500
volt
DC
reference
supply.
Meter
cannot
be
brought
to
zero
with
CALIBRATE
control.
Measurements
are
out
of
tolerance
on
one
range
other
than
the
500
volt
range.
Measurements
are
out
of
tolerance
on
all
ranges
other
than
the
500
volt
range.
4-2
V104,
У105,
or
V106
defective.
One
of
the
sampling
string
resistors
(R119,
R120,
R121)
is
changing
value
rapidly
as
the
instrument
warms
up.
Drifting
standard
cell
EMF
possibly
caused
by
previous
shorting
or
inverting
of
the
cell
(825A
and
825AR
only).
Excessive
aging
of
V105.
Out
of
calibration.
One
or
more
resistors
in
500
volt
DC
sampling
string
has
shifted
in
value.
Standard
cell
EMF
has
shifted
(825A
and
825AR
only).
A
reference
supply
output
range
resistor
on
the
range
resistor
printed
circuit
board
is
out
of
tolerance.
R320
or
R321
has
shifted
in
value.
Figure
4-2.
Check
V104,
V105,
and
V106
by
replacement.
Locate
faulty
resistor
by
heating
slightly
with
a
soldering
iron
held
near
the
resistor
while
ob-
serving
the
500
volt
DC
calibration.
Allow
time
for
the
standard
cell
to
stabilize.
Several
hours
should
be
sufficient.
Check
V104
by
replacement.
Recalibrate
per
paragraph
4-4.
Recalibrate
per
paragraph
4-4
and
observe
sta-
bility
for
48
hours.
If
500
volt
DC
reference
supply
remains
stable
replacement
of
resistor
is
unnecessary.
Recalibrate
instrument
and
observe
Stability
of
reference
supply.
Replacement
of
the
standard
cell
may
be
necessary.
If
the
trouble
occurs
in
the
50
volt
range,
R324
has
shifted;
in
the
5
volt
range,
R325
or
R327
has
shifted;
in
the
0.5
volt
range,
R328
or
R330
has
shifted.
It
may
be
possible
to
correct
by
recalibration.
If
not,
replace
faulty
resistor.
It
may
be
possible
to
correct
all
range
voltages
by
recalibration.
If
not,
replace
faulty
resistor.
TROUBLESHOOTING
(Sheet
1
of
2)

Related product manuals