8505A
Mathematic Operations;
To store the displayed value
as an
offset orscaling
value and enter the respective
mode:
STORE
('
OFFSET
\
SCALING
To store
a
numeric
entry as an offset or scaling value and enter
the respective
mode:
STORE (numerics)
OFFSET
SCALING
Special Operations:
To store the
displayed value
as a
high
or low limit and enter the
L, 3
mode:
STORE
To store
a
numeric
entry as a high or low
limit
and enter the Limits mode:
STORE (numerics)
MANUAL/
EXTERNAL
CONTROL
RESET LDL/RMT
The multimeter utilizes fourteen dual purpose push buttons
as
numeric
entry keys. When
programming sample
settings or filter modes, a one- or
two-digit
entry is necessary. When programming limits, orscaling and offset values,
a
longer entry may
be
required. Therefore, the following
sequence is available:
Mantissa Field: enter a maximum
of seven digits, with decimal point. Toggle polarity
(±)
button
as required.
Exponent Field: push EXP button,
then enter a single digit exponent. Toggle the polarity button
again to set the exponent polarity.
Use the CE push button at any time prior
to the terminator
(last
button in a store numeric sequence)
to
clear
all numerics and revert to a
“?”
display. Fresh numerics
may then be entered immediately.
Figure
2-2.
Front
Panel Push Buttons (cant)
2-13