EasyManua.ls Logo

Fluke 863 - Rel Test;Touch Hold Test

Fluke 863
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
867B,863
Service Manual
4-20
Rel Test/Touch Hold Test 4-17.
1. Turn on the GMM to VDC, and apply 1000V dc.
2. Press t[Rel].
3. Check for a Rel reading between the limits shown in Table 4-19.
Table 4-19. Rel Performance Test
Input Input Source Lower Limit Upper Limit
1000.0V dc 5700A using GMM volts input -0.2 0.2
4. Press é.
5. On the 5700A, press l to select Standby.
6. Check for a GMM reading within the limits shown in Table 4-20.
Table 4-20. Touch Hold Performance Test
Input Input Source Lower Limit Upper Limit
0V dc (open) -0.2 0.2

Table of Contents

Other manuals for Fluke 863

Related product manuals