Remote Operation 
 
Error Codes 
 
  101 
 
Table 35. Calibration Errors 
Error Code  Error Text  Description 
701  CAL Secure Code Case Invalid characters in calibration secure code. The secure code 
must contain only characters 0-9 and A-Z. No lower case or 
other special characters are allowed. 
702  CAL Secure  Cal secured. 
703  CAL Secure code  Invalid secure code. 
704  CAL Secure code too 
long 
Secure code too long. 
706  CAL Invalid step 
number 
Invalid calibration step number. 
707  CAL Reference  Cal reference value out of tolerance. 
708  CAL Not calibrating  CAL? only works if you are calibrating. 
709  CAL Abort  Calibration aborted. 
710 CAL Sequence Calibration steps out-of-sequence. 
711  CAL No measurement  No measurements taken during calibration. 
712  CAL Overload  Overload at input during calibration. 
713  CAL Too far  Calibration measurements out-of-tolerance. 
714  CAL Math  Math error during calibration. 
715  CAL Timeout  Timeout error during calibration.