Operational Differences 
When changing the measurement resolution, the  setting is not an available 
selection. This selection would result in a measurement rate that would be much faster 
than the 8842A could achieve. 
The 8842A emulation selection does not appear when attempting to select a command 
language whenever the 8845A or 8846A has a port selection of RS-232 or LAN. This is 
due to the fact that the Fluke 8842A does not have a serial or Ethernet interface. This also 
means that RS-232 or LAN port selections are not available whenever the 8845A or 
8846A is set to the 8842A emulation mode. 
Pressing RESET ( followed by ) resets the Meter back to 8842A defaults.