Remote Operation
Supported SCPI Commands
41
Notes [1] For ac measurements, resolution is fixed at 6½ digits. The resolution
parameter only affects the front panel display.
[2] For frequency measurements, the Meter uses one range for all inputs
between 3 Hz, and 300 kHz. With no input signal applied, frequency
measurements return zero.
[3] For period measurements, the Meter uses one range for all inputs
between 0.33 seconds, and 3.3 µsec. With no input signal applied,
period measurements return zero.
[4] The range and resolution are fixed for temperature measurements.
However, placing PT100_385 or PT100_392 into the range field will
set the RTD type. Placing CUST1 into the range field will cause the
Meter to use the values set into the R
0
and Alpha parameters.
[5] The range and resolution are fixed for diode test: 10 Vdc range and 5½
digits. When set to ON, the low current parameter sets the diode test
current to 0.1 mA. When OFF, the current is 1 mA. When set to ON,
the high voltage parameter sets the diode test voltage to 10 volts. When
OFF, the voltage is 5 volts. To specify a diode test voltage, you must
also include ON or OFF for the low current parameter.
[6] The range and resolution are fixed for continuity tests: 1 kΩ range and
5½ digits.
[7] DC measurement voltage divided by dc reference voltage. For dc ratio,
connect the reference HI/LO to the HI/LO sense terminals of the Meter
and the measurement voltage to the HI/LO input terminals of the
Meter. Note the specified measurement range applies to the input
terminals only.
Note
For best results in RATIO, the two input commons must be shorted at the
Meter terminals. The analog filter (AFLTR) should be off.
Table 18. Preset Conditions for the MEASure? and CONFigure Commands
Parameter Setting
AC Filter (DET:BAND) 20 Hz
Autozero (ZERO:AUTO) OFF if NPLC <1, ON if NPLC >=1
DC Digital Filter ON
Input Impedance (INP:IMP:AUTO) OFF
Samples per trigger (SAMP:COUN) 1
Trigger count (TRIG:COUN) 1
Trigger delay (TRIG:DEL) Automatic
Trigger source (TRIG:SOUR) Internal (immediate)
Math Function (CALCulate subsystem) OFF