8845A/8846A
Programmers Manual
52
CALCulate:FUNCtion(?) {NULL|DB|DBM|AVERage|LIMit}
Description Selects one of five math functions. This setting is stored in volatile
memory. Selecting a math function disables any math function which may
have been set previously. For the selected math function to take effect, the
CALCulate:STATe must be ON.
Parameters <n> = NULL Uses the present reading as an offset to zero
the returned reading.
DB Selects the DB mode.
DBM Selects the dBm mode.
AVERage Selects the average mode
LIMit Selects the Limit Testing mode.
Examples CALC:FUNC DBM Selects the dBm mode.
Query CALC:FUNC? Returns the selected math function.
CALCulate:STATe(?) <n>
Description Activates or deactivates the math function selected with CALC:FUNC
command. This setting is stored in volatile memory.
Parameters <n> = ON or 1 Activates the math function set previously by
CALC:FUNC command.
OFF or 0 Deactivates the math function.
Examples CALC:STAT ON Turns on the math function.
Query CALC:STAT? Returns the state of the math function. (0 = OFF
and 1 = ON)
CALCulate:AVERage:MINimum?
Description Retrieves the minimum value measured during a statistics operation. This
value is cleared whenever the Meter is powered off then back on, a
statistics function is enabled, or the Meter receives a reset through the
remote interface. This setting is stored in volatile memory.
Query CALC:AVER:MIN? Returns the minimum value of the statistics operation.
CALCulate:AVERage:MAXimum?
Description Retrieves the maximum value measured during a statistics operation. This
value is cleared whenever the Meter is powered off then back on, a
statistics function is enabled, or the Meter receives a reset through the
remote interface. This setting is stored in volatile memory.
Query CALC:AVER:MAX? Returns the maximum value of the statistics operation.
CALCulate:AVERage:AVERage?
Description Retrieves the average value measured during a statistics operation. This
value is cleared whenever the Meter is powered off then back on, a
statistics function is enabled, or the Meter receives a reset through the
remote interface. This setting is stored in volatile memory.
Query CALC:AVER:AVER? Returns the average value of the statistics operation.