8845A/8846A
Programmers Manual
70
Function Commands and Queries
The commands in Table 21 lists the measurement function commands for the Fluke 45.
See Tables 22A and 22B for ranges and measurement rates. Commands under Primary
Display and Secondary Display cause functions to be performed using the primary
display or secondary display, respectively.
Table 21. Fluke 45 Function Commands and Queries
Commands
Primary
Display
Secondary
Display
Function
AAC
ADC
CONT
DIODE
FREQ
FUNC1 ?
OHMS
VAC
VDC
AAC2
ADC2
CLR2
DIODE2
FREQ2
FUNC2?
OHMS2
VAC2
VDC2
AC current
DC current
Clears measurement from secondary display if one shown.
Continuity test. Available in the primary display only.
Diode test
Frequency
Meter returns the function selected for the primary display as
command mnemonic.
EXAMPLE: If frequency is selected for the primary display,
"FUNC1?" returns "FREQ".
Meter returns the function selected for the secondary display as
command mnemonic. If the secondary display is not in use, an
Execution Error is generated.
EXAMPLE: If frequency is selected for the secondary display,
FUNC2? returns "FREQ".
Resistance
AC volts
DC volts