Performance Test and Calibration
Performance Tests 3
3-9
Table 3-4. 8846A AC Volts Verification Steps
Nominal Input 90-day Test Limits 1-year Test Limits
Ampl. Freq.
Range
High Low High Low
100.0 mV
[1]
10 Hz 0.100 100.09 mV 99.91 mV 100.1 mV 99.9 mV
100.0 mV 20 kHz 0.100 100.09 mV 99.91 mV 100.1 mV 99.9 mV
100.0 mV 50 kHz 0.100 100.16 mV 99.84 mV 100.17 mV 99.83 mV
100.0 mV 100 kHz 0.100 100.68 mV 99.32 mV 100.68 mV 99.32 mV
100.0 mV 300 kHz 0.100 104.5 mV 95.5 mV 104.5 mV 95.5 mV
1 V
[1]
10 Hz 1 1.0008 V 999.2 mV 1.0009 V 999.1 mV
1 V 20 kHz 1 1.0008 V 999.2 mV 1.0009 V 999.1 mV
1 V 50 kHz 1 1.0016 V 998.4 mV 1.0017 V 998.3 mV
1 V 100 kHz 1 1.0068 V 993.2 mV 1.0068 V 993.2 mV
1 V 300 kHz 1 1.045 V 955.0 mV 1.045 V 955.0 mV
10 V
[1]
10 Hz 10 10.008 V 9.992 V 10.009 V 9.991 V
10 V
[1]
20 kHz 10 10.008 V 9.992 V 10.009 V 9.991 V
10 V 50 kHz 10 10.016 V 9.984 V 10.017 V 9.983 V
10 V 100 kHz 10 10.068 V 9.932 V 10.068 V 9.932 V
3 V 300 kHz 10 3.17 V 2.83 V 3.17 V 2.83 V
100 V 45 Hz 100 100.08 V 99.92 V 100.09 V 99.91 V
100 V
[1]
20 kHz 100 100.08 V 99.92 V 100.09 V 99.91 V
100 V 50 kHz 100 100.16 V 99.84 V 100.17 V 99.83 V
100 V
[1]
100 kHz 100 100.68 V 99.32 V 100.68 V 99.32 V
1000 V 45 Hz 1000 1000.800 999.200 1.0009 kV 999.1 V
1000 V 1 kHz 1000 1000.800 999.200 1.0009 kV 999.1 V
1000 V
[1]
10 kHz 1000 1000.800 999.200 1.0009 kV 999.1 V
320 V 20 kHz 1000 320.460 319.540 320.492 V 319.508 V
320 V 50 kHz 1000 320.852 V 319.148 V 320.884 V 319.116 V
320 V 100 kHz 1000 322.72 V 317.28 V 322.72 V 317.28 V
[1] 5520A must be used with 8508A to obtain suitable test uncertainty ratio.