8845A/8846A
Calibration Manual
3-10
Table 3-5. 8845A AC Volts Verification Steps
Nominal Input 90-day Test Limits 1-year Test Limits
Ampl. Freq.
Range
High Low High Low
100.0 mV
[1]
10 Hz 0.100 100.09 mV 99.91 mV 100.1 mV 99.9 mV
100.0 mV 20 kHz 0.100 100.09 mV 99.91 mV 100.1 mV 99.9 mV
100.0 mV 50 kHz 0.100 100.16 mV 99.84 mV 100.17 mV 99.83 mV
100.0 mV 100 kHz 0.100 100.68 mV 99.32 mV 100.68 mV 99.32 mV
100.0 mV 300 kHz 0.100 104.5 mV 95.5 mV 104.5 mV 95.5 mV
1 V
[1]
10 Hz 1 1.0008 V 999.2 mV 1.0009 V 999.1 mV
1 V 20 kHz 1 1.0008 V 999.2 mV 1.0009 V 999.1 mV
1 V 50 kHz 1 1.0016 V 998.4 mV 1.0017 V 998.3 mV
1 V 100 kHz 1 1.0068 V 993.2 mV 1.0068 V 993.2 mV
1 V 300 kHz 1 1.045 V 955.0 mV 1.045 V 955.0 mV
10 V
[1]
10 Hz 10 10.008 V 9.992 V 10.009 V 9.991 V
10 V
[1]
20 kHz 10 10.008 V 9.992 V 10.009 V 9.991 V
10 V 50 kHz 10 10.016 V 9.984 V 10.017 V 9.983 V
10 V 100 kHz 10 10.068 V 9.932 V 10.068 V 9.932 V
3 V 300 kHz 10 3.17 V 2.83 V 3.17 V 2.83 V
100 V 45 Hz 100 100.08 V 99.92 V 100.09 V 99.91 V
100 V
[1]
20 kHz 100 100.08 V 99.92 V 100.09 V 99.91 V
100 V 50 kHz 100 100.16 V 99.84 V 100.17 V 99.83 V
100 V
[1]
100 kHz 100 100.68 V 99.32 V 100.68 V 99.32 V
750 V 45 Hz 750 750.600 749.400 750.675 V 749.325 V
750 V 1 kHz 750 750.600 749.400 750.675 V 749.325 V
750 V 10 kHz 750 750.600 749.400 750.675 V 749.325 V
320 V 20 kHz 750 320.385 319.615 320.417 V 319.583 V
320 V 50 kHz 750 320.727 V 319.273 V 320.759 V 319.241 V
320 V 100 kHz 750 322.52 V 317.48 V 322.52 V 317.48 V
[1] 5520A must be used with 8508A to obtain suitable test uncertainty ratio.