Performance Test and Calibration
Performance Tests 3
3-11
Table 3-6. 8845A/8846A AC Volts Frequency Verification Steps
Nominal Input 90-day Test Limits 1-year Test Limits
Ampl. Freq. High Low High Low
1 V 10 Hz 10.00300 Hz 9.99700 Hz 10.00300 Hz 9.99700 Hz
1 V 40 Hz 40.0040 Hz 39.9960 Hz 40.01200 Hz 39.98800 Hz
100 mV 300 kHz 300.030 kHz 29.99700 kHz 300.0300 kHz 29.99700 kHz
100 mV 1 MHz
[1]
1.000100 MHz 999.9000 kHz 1.0001000 MHz 999.9000 kHz
[1] 8846A Only.