Appendices
Verification Forms A
Table A-1. Blank Verification Record for 90-Day Specifications (cont)
Nominal Output Test Limits
Ampl. Freq.
Range
High Low
Results Pass/Fail
8845A/8846A 2-Wire Resistance
0 100 0.0040 0
100 100 100.0120 99.9880
0 1000 10.0E-3 0
1000 1000 1000.090 999.910
0 10000 100.0E-3 0
10000 10000 10000.90 9999.10
0 100000 1.000000 0
100000 100000 100009.0 99991.0
0 1000000 10.00000 0
1000000 1000000 1000090 999910
0 10000000 100.0000 0
10000000 10000000 10002100 9997900
0 100000000 10000.00 0
100000000 100000000 100810000 99190000
8846A Only 2-Wire Resistance
0 1000000000 100000.0 0
1000000000 1000000000 1015100000 984900000
8845A/8846A Optional 2X4 Test Lead (Split Lead) test
0 100 4.0E-3 0
100 100 100.0120 99.9880
8846A Optional Rear Panel Test
10 V DCV (10V) 10.00023 V 9.99977 V
1000 4-W (1 k) 1000.09 999.91
100 mA DCI (100 mA) 0.100035 A 0.099965 A
8845A Optional Rear Panel Test
10 V DCV (10V) 10.00025 V 9.99975 V
1000 4-W (1 k) 1000.09 999.91
100 mA DCI (100 mA) 0.100035 A 0.099965 A