8845A/8846A
Calibration Manual
A-6
Table A-1. Blank Verification Record for 90-Day Specifications (cont)
Nominal Output Test Limits
Ampl. Freq.
Range
High Low
Results Pass/Fail
8846A Capacitance
0 1.0E-9 25.0E-12 0
1.0E-9 1.0E-9 1.045E-9 955.0E-12
10.0E-9 10.0E-9 10.15E-9 9.85E-9
100.0E-9 100.0E-9 101.5E-9 98.5E-9
1.0E-6 1.0E-6 1.015E-6 985.0E-9
10.0E-6 10.0E-6 10.15E-6 9.85E-6
100.0E-6 100.0E-6 101.5E-6 98.5E-6
1.0E-3 1.0E-3 1.015E-3 985.0E-6
10.0E-3 10.0E-3 10.15E-3 9.85E-3
100.0E-3 100.0E-3 104.2E-3 95.8E-3
8846A DC Current
0 100.0E-6 25.0E-9 -25.0E-9
100.0E-6 100.0E-6 100.065E-6 99.935E-6
-100.0E-6 100.0E-6 -99.935E-6 -100.065E-6
0 1.0E-3 50.0E-9 -50.0E-9
1.0E-3 1.0E-3 1.00035E-3 999.65E-6
-1.0E-3 1.0E-3 -999.65E-6 -1.00035E-3
0 10.0E-3 2.0E-6 -2.0E-6
10.0E-3 10.0E-3 10.005E-3 9.995E-3
-10.0E-3 10.0E-3 -9.995E-3 -10.005E-3
0 100.0E-3 5.0E-6 -5.0E-6
100.0E-3 100.0E-3 100.035E-3 99.965E-3
-100.0E-3 100.0E-3 -99.965E-3 -100.035E-3
0 1 200.0E-6 -200.0E-6
1 1 1.0006E+0 999.4E-3
-1 1 -999.4E-3 -1.0006E+0
0 3 600.0E-6 -600.0E-6
1.9 3 1.90212E+0 1.89788E+0
-1.9 3 -1.89788E+0 -1.90212E+0