Fluke 9140 - 7.4 Temperature Display Hold

Fluke 9140
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the same hole for the reference thermometer and the test probe may have
better results. This however requires switching the probes which takes more
time. One must allow a few minutes after inserting the probes for the tempera
-
ture to stabilize before making measurements. Because of temperature varia
-
tions along the length of the well, best results are obtained when comparing
probes of similar construction and inserting them the same depth into the well.
19
6 General Operation
Calibrating Probes