b. Press SET then use the UP or DOWN keys until the correct numerical
setting is displayed. Press SET to accept the new value.
c. Repeat step b. for ALPHA and DELTA.
10.2.3 Accuracy & Repeatability
1. Check the accuracy of the dry-well at various points over the calibration
range.
2. If dry-well does not pass specification at all set-points, repeat Calibration
Procedure.
45
10 Calibration Procedure
Calibration Procedure