ERRATA tl
On page 4-2, add the following sentence to the’end of paragraph 4-B:
The.752A must be returned to the factory for this service.
On page 4-11, following paragraph 4-47,
ADD :
NOTE
Problems involving shifts in resistor
values or unstable resistor values that
can not be compensated, require the 752A
to be returned to the factory for repair.
This is due to the matching requirements
for all four modules and the Bridge and
Compensation board.
On pages 5-3 and 5-4, Table 5-l:
CHANGE: AllRESISTOR MODULES, MATCHED ASSEMBLY
IModuIe Exchange Recommended11
TO:
AIIRESISTOR MODULES, MATCHED ASSEMBLY
IFactory Maintenance Recommended11
DELETE:
MP421....
ERRATA #2
On page 1-2, Table
1-2,
under the PONER COEFFICIENT EFFECT ON RATIO***,
CHANGE: 1O:l Ratio .............
CO.05 ppm of input @ 1OOV
TO: 1O:l Ratio .............
(0.05 ppm of output @ IOOV
CHANGE: 1OO:l Ratio . . . . . . . . . . . .
CO.3
ppm of input @ 1OOOV
TO: 100: 1 Ratio . . . . . . . . . . .‘.
10.3
ppm of,output,@ IOOOV
ERRATA
#3
On page 2-4, add the following note prior to paragraph 4-22:
NOTE
To minimize noise effects in 0.1 and 1V settings, the
Null Detector is configured in opposite polarity.