English Instruction SheetPanel 1
®
707
Loop Calibrator
Instruction Sheet
Introduction
The Fluke 707 Loop Calibrator (hereafter, the Calibrator) is a
compact sourcing and measuring tool. The Calibrator tests current
loops of 0-20 mA or 4-20 mA and measures dc voltage to 28 V. It
comes with a set of alligator-clip test leads, a 9 V alkaline battery,
and this Instruction Sheet.
The Calibrator is a IEC 61010, CAT I 30 V, Pollution Degree 2
instrument. A CAT I instrument is designed to protect against
transients from high-voltage, low-energy sources, like electronic
circuits or a copy machine, for example.
Calibrator Capabilities
Function Range Resolution
Measure V dc
28 V 1 mV
Measure mA dc
Source mA dc
0 to 24 mA 1 μA
Source loop power
24 V dc N/A
Battery Saver
The Calibrator automatically turns off after 30 minutes of inactivity.
To reduce this time or disable this feature:
1. With the Calibrator OFF, press D.
PSXX is displayed, where XX is the turn-off time in minutes.
OFF means the power saver is disabled.
2. Turn m to decrease or increase the turn-off time.
To disable, turn m until the display shows OFF.
3. The Calibrator resumes normal operation after 2 seconds.
August 2001 (English) Rev. 2, 2/08
© 2001-2008 Fluke Corporation. Product specifications are subject to
change without notice. All rights reserved.

Other manuals for Fluke 707

Question and Answer IconNeed help?

Do you have a question about the Fluke 707 and is the answer not in the manual?

Fluke 707 Specifications

General IconGeneral
Auto stepYes
Control typeButtons
Loop voltage30 V
Storage temperature (T-T)-30 - 60 °C
Operating temperature (T-T)-10 - 55 °C
Voltage measurement accuracy0.015 %
Voltage measurement resolution0.001 V
Operating relative humidity (H-H)0 - 90 %
Battery voltage9 V
Battery technologyAlkaline
Weight and Dimensions IconWeight and Dimensions
Depth41 mm
Width70 mm
Height152 mm
Weight224 g

Summary

Safety Warnings and Symbols

General Warnings and Cautions

Provides essential safety advice to avoid electric shock, injury, or damage to the Calibrator.

mA Sourcing Modes

Using mA Sourcing

Guides on using the Calibrator to output current for testing 0-20 mA and 4-20 mA loops.

Simulating Transmitter Operation

Simulate mA Operation

Explains how to simulate a 2-wire transmitter by regulating loop current with a 12V to 28V loop supply.

Advanced mA Output Functions

Measuring dc mA

Measuring dc mA Procedure

Step-by-step guide to measure direct current (mA) using the Calibrator.

Measuring mA with Loop Power

Loop Power Measurement Procedure

Instructions for powering a transmitter and reading loop current simultaneously using loop power.

Measuring dc Volts

Measuring dc Volts Procedure

Step-by-step guide to measure direct current voltage (V dc) using the Calibrator.

Maintenance and Troubleshooting

General Maintenance Warnings

Advises on safety precautions during servicing and maintenance of the Calibrator.