Fluke 707 - Measuring dc Volts; Measuring dc Volts Procedure

Fluke 707
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English Instruction SheetPanel 10
Measuring dc Volts
W Caution
To prevent damage to the unit under test, ensure that the
Calibrator is in the correct mode before connecting the
test leads.
To measure dc Volts:
1. Press A to step to MEASURE mode.
MEASURE V is displayed.
2. Touch test lead probes across the load or power source.
30 V dc max.
COM
+
707
LOOP CALIBRATOR
30V
CAT
Press until MEASURE
V dc is displayed.