Port for Fluke 700 Series Pressure Module. Supplied with 3/8-in. BSP to 1/4-in. NPT adapter.
Push in from this side for vacuum, and from the other side for pressure. Use a small screwdriver.
Flexible hose to pressure instrument under test.
1/4-in. NPT adapter.
Tighten nuts to limit pressure, shortening the compression stroke.
Unscrew to vent pressure. Tighten valve to close before applying pressure.
Allows precise adjustment of applied pressure.
Attach 700 Series Pressure Module to master instrument port.
Connect pressure module to the calibrator as shown.
Connect test pressure port hose to the instrument under test.
Ensure the pressure/vacuum button is in the desired position.
Unscrew the pressure release valve to vent pressure from the pump.
Zero the pressure module according to its specific procedure.
Turn the fine adjustment knob to mid-range initially.
Port for Fluke 700 Series Pressure Module. Supplied with 3/8-in. BSP to 1/4-in. NPT adapter.
Push in from this side for vacuum, and from the other side for pressure. Use a small screwdriver.
Flexible hose to pressure instrument under test.
1/4-in. NPT adapter.
Tighten nuts to limit pressure, shortening the compression stroke.
Unscrew to vent pressure. Tighten valve to close before applying pressure.
Allows precise adjustment of applied pressure.
Attach 700 Series Pressure Module to master instrument port.
Connect pressure module to the calibrator as shown.
Connect test pressure port hose to the instrument under test.
Ensure the pressure/vacuum button is in the desired position.
Unscrew the pressure release valve to vent pressure from the pump.
Zero the pressure module according to its specific procedure.
Turn the fine adjustment knob to mid-range initially.
| Pressure Media | Non-corrosive gases |
|---|---|
| Pressure Connection | 1/4 in NPT female |
| Materials | Anodized aluminum and stainless steel |
| Resolution | 0.01 psi |
| Power Supply | Not applicable |
| Operating Temperature | 32 to 122°F (0 to 50°C) |
| Storage Temperature | -20°C to 60°C |
| Accuracy | ±0.05% full scale |