English Instruction SheetPanel 15
Accuracy Specifications
Accuracy is specified for 1 year after calibration at operating
temperatures of 18 °C to +28 °C and is given as:
±( [ % of reading ]+[ counts ] )
MEASURE V dc
Range: +28 V (+30 V max)
Resolution: 1 mV
Input Impedance: 1 MΩ
Accuracy: ±(0.015 % of reading+2 counts)
MEASURE mA dc
Range: 20 mA (24 mA max)
Resolution: 1 μA
Accuracy: ±(0.015 % of reading+2 counts)
SOURCE / SIMULATE mA dc
Range: 0 mA to 20 mA (24 mA max)
Resolution: 1 μA
Accuracy: ±(0.015 % of reading+2 counts)
Source mode:
Compliance: To 1200 Ω at 20 mA
To 950 Ω at 20 mA in HART
TM
Mode
Simulate mode:
External loop voltage requirement: 24 V nominal, 30 V maximum,
12 V minimum
Loop Power
24 V
Percent display
-25 % to 125 %
Input / Output Protection
Fuseless protection