COMMAND REFERENCE 4 - 87
SENSe:VOLTage<n>[:DC]:RANGe:OFFSet
Syntax:
SENSe:VOLTage<n>[:DC]:RANGe:OFFSet <NRf>
| MINimum | MAXimum
<n> [1] | 2 | 3 | 4
Note: Channel 3 and 4 not applicable for PM33x0B.
<NRf> The vertical offset for channel <n> in volts.
MINimum Selects the minimum possible vertical offset.
MAXimum Selects the maximum possible vertical offset.
Query form:
SENSe:VOLTage<n>[:DC]:RANGe:OFFSet?
[ MINimum | MAXimum]
Response:
<NR3>
<NR3> The vertical offset for channel <n> in volts.
If MINimum was specified, the minimum possible value is returned.
If MAXimum was specified, the maximum possible value is
returned.
Description:
Controls the vertical offset for an input channel. The vertical offset for channel <n>
is expressed in volts. If a detectable probe is attached, the offset value is
considered to be at the probe tip, otherwise at the BNC plug.
After a
*
RST command, the vertical offset for each channel is zero.
Coupled values:
The range of the offset value is directly coupled to the range of the vertical
sensitivity per division (SENSe:VOLTage<n>:RANGe:PTPeak).
Example:
Send
SENSe:VOLTage2:RANGe:OFFSet 1E-1 Sets 100 mV offset for
channel 2.
Front panel compliance:
The SENSe:VOLTage<n>:RANGe:OFFSet command is the remote equivalent of
the front panel POS knobs.