4 - 98 COMMAND REFERENCE
SYSTem:COMMunicate:SERial[:RECeive]:BAUD
SYSTem:COMMunicate:SERial:TRANsmit:BAUD
SYSTem:COMMunicate:SERial[:RECeive]:BITS
SYSTem:COMMunicate:SERial:TRANsmit:BITS
SYSTem:COMMunicate:SERial[:RECeive]:PACE
SYSTem:COMMunicate:SERial:TRANsmit:PACE
SYSTem:COMMunicate:SERial[:RECeive]:PARity[:TYPE]
SYSTem:COMMunicate:SERial:TRANsmit:PARity[:TYPE]
Syntax:
SYSTem:COMMunicate:SERial[:RECeive]:BAUD <baudrate>
SYSTem:COMMunicate:SERial:TRANsmit:BAUD <baudrate>
<baudrate> 75 | 110 | 150 | 300 | 600 | 1200 | 2400 | 4800 | 9600
| 19200 | 38400 | MIN | MAX
SYSTem:COMMunicate:SERial[:RECeive]:BITS 7 | 8
SYSTem:COMMunicate:SERial:TRANsmit:BITS 7 | 8
SYSTem:COMMunicate:SERial[:RECeive]:PACE XON | NONE
SYSTem:COMMunicate:SERial:TRANsmit:PACE XON | NONE
XON Enables the X-on/X-off handshake.
NONE Disables the X-on/X-off handshake.
SYSTem:COMMunicate:SERial[:RECeive]:PARity[:TYPE]
EVEN | ODD | NONE
SYSTem:COMMunicate:SERial:TRANsmit:PARity[:TYPE]
EVEN | ODD | NONE
Query form:
SYSTem:COMMunicate:SERial[:RECeive]:BAUD? [MIN | MAX]
SYSTem:COMMunicate:SERial:TRANsmit:BAUD? [MIN | MAX]
Response:
75 | 110 | 150 | 300 | 600 | 1200 | 2400 | 4800 | 9600 | 19200 | 38400
If MINimum was specified, 75 is returned.
If MAXimum was specified, 38400 is returned.
Query form:
SYSTem:COMMunicate:SERial[:RECeive]:BITS?
SYSTem:COMMunicate:SERial:TRANsmit:BITS?
Response:
7 | 8