Introduction
The i30s AC/DC Current Clamp has been designed for use with
oscilloscopes and DMMs for accurate, non-intrusive
measurement of both AC, DC, and complex waveform currents.
The i30 AC/DC Current Clamp has been designed for use with
DMMs only. Using advanced Hall Effect technology the i30s can
accurately measure currents with a resolution of 1 mA from 5 mA
to 30 A over the frequency range of DC to 100 kHz. These
features make it a powerful tool for use in inverters, switch mode
power supplies, industrial controllers and other applications
requiring current measurements and/or waveform analysis.
Symbols
The table below lists the symbols used on the product and/or in
this manual.
Symbol Description
~
Do not dispose of this product as unsorted municipal
waste. Contact Fluke or a qualified recycler for
disposal.
W
Important Information. See manual.
T
Double insulation.
-
Do not apply around or remove from the HAZARDOUS
LIVE conductors.
)
Conforms to Canadian Standards Association.
P
Complies with the relevant European standards.
2 / 11