Page 3
Use of this equipment in a manner not specified
herein may impair the protection provided by
the equipment.
CAT III equipment is designed to protect against
the transients in the equipment in fixed
equipment installations, such as distribution
panels, feeders and short branch circuits, and
the lighting systems in large buildings.
Specifications
Electrical Characteristics
All accuracies stated at 23°C ± 1°C
Current range 20 A AC
RMS
or DC
Measuring range ± 30 A
Overload capacity 500 A
Output sensitivity 100 mV/A
Accuracy (at 25 °C) ± 1 % of reading ± 2 mA
Resolution ± 1 mA
Typical output noise
level
200 µV
RMS
Load impedance > 100 k Ohms 100 pF
Conductor position
sensitivity
± 1 % relative to center reading
Frequency range
i30s DC to 100 kHz (0.5 dB)
i30 DC to 20 kHz (0.5 dB)
di / dt following 20 A / µs
Response time Better than 1 µs
Temperature
coefficient
± 0.01 % of reading / °C
Operating humidity: 15 to 85 % (non
condensing)C
Altitude: 2000 m
Power supply 9 V Alkaline, NEDA 1604/PP3
IEC 6LR61
Working voltage
(see Safety
Standards)
300 V AC
RMS
or DC
Battery life 30 hours, low battery indicator