Maintenance
PERFORMANCE
TESTS
Resistance
Test
3-10.
1. Select
the ohms
function
on the UUT.
2.
Connect the
Ohms
Calibrator or
Decade
Resistor
to
the
vn-w-
and COM input
terminals
of the
UUT.
3. Referring
to Table
3-4,
set
the
Ohms Calibrator or
Decade Resistor
to
the
resistance
value
indicated in
steps
1
through 6. Verify
that the display
reading
is
within
the
limits
shown.
Table
3-4.
Resistance
Test
STEP
INPUT
RESISTANCE
DISPLAY
READING
1
short
0.0
to
0.2Q
Decades
of 1:
2
100Q
99.3 to
100.7Q
(plus 0 reading)
3
1
kQ
994
to
1006
Q
4
10 kQ
9.994 to 10.06
kQ
5
100
kQ
99.94 to 100.6
kQ
6
1
MQ
.994 to 1.006
MQ
7
10
MQ
9.79 to 10.21
MQ
Decades of
1.9:
2 i9on
188.8 to
191.2Q
(plus 0 reading)
3
1.9 kQ
1889 to 1911 Q
4
19 kQ
18.89
to 19.11
kQ
5
190 kQ
188.9 to
191.1 kQ
6
1.9 MQ
1.889 to
1.911 MQ
7 19
MQ
18.61
to 19.39 MQ
Decades of
2.7:
2 270Q
268.4
to 271.
6Q
(plus
0
reading)
3
2.7
kQ
2685 to
271 5Q
4
27 kQ
26.85 to
27.15 kQ
5
270
kQ
268.5
to 271.5 kQ
6
2.7
MQ
2.685
to 2.715 MQ
7
27 MQ
26.45 to 27.55
MQ
Diode
Test
3-1
1
.
To
test
the
Series II
meters, perform the following
steps:
1.
Set
the
UUT
to the Diode Test function.
NOTE
On Fluke
5100
series
calibrators,
activate
the
50Q.
divider override.
2.
Apply an input
from the DC Voltage
Calibrator
of
+.090V
dc to the vn-w- and
COM input
terminals of the UUT,
and verify
that the
beeper is
on.
3. Increase
the DC
Voltage
Calibrator
output
to
+.1 10V
dc,
and verify
that
the
beeper
is off.
3-8