Digital Multimeter 
  General Specifications   
 13 
General Specifications 
Accuracy is specified for 1 yr after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy 
specifications take the form of ±([ % of Reading ] + [ Counts ]). 
Maximum voltage between any  
terminal and earth ground ............................................ 1000 V 
Fuse protection for mA inputs ..................................... 0.44 A, 1000 V, IR 10 kA 
Fuse protection for A inputs ........................................ 11 A, 1000 V, IR 17 kA FAST Fuse 
Display 
Digital .............................................................................. 6000 counts, updates 4/sec 
Bar Graph ........................................................................ 33 segments; Updates 32/sec 
Frequency ....................................................................... 10,000 counts 
Capacitance .................................................................... 1,000 counts 
Altitude 
Operating .................................................................... 2,000 meters 
Storage ........................................................................ 12,000  meters 
Temperature 
Operating .................................................................... -10 °C to +50 °C 
Storage ........................................................................ -40 °C to +60 °C 
Temperature coefficient ................................................ 0.1 X (specified accuracy) / °C (< 18 °C or > 28 °C) 
Safety 
General ....................................................................... IEC 61010-1: Pollution Degree 2 
Measurement .............................................................. IEC 61010-2-033: CAT IV 600 V / CAT III 1000 V
 
Relative Humidity (Maximum Non-condensing) ......... 90 % to 35 °C 
75 % to 40 °C  
45 % to 50 °C