EasyManua.ls Logo

Fluke PM6681 - Page 14

Fluke PM6681
153 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EXT ARM
input
-
Select
non AUTO.
-
Settings for pulse generator: single
shot pulse,
amplitude
TTL
=
0-2
Vpp, and duration
=
10
ns.
-
Connect
a pulse generator to EXT
ARM input.
-
Press START ARM key.
-
Press
SELECT
key until display
shows ’POS’, confirm
with ENTER key three times.
-
The
counter
does not measure.
-
Apply
one single pulse
to
EXT
ARM input.
-
The counter measures once and
shows
10 MHz on the
display.
TRIG LEVEL A&B Outputs
-
Press the PRESET key, to set the
timer/counter in the
default
setting.
-
Connect a voltmeter to TRIG
LEVEL A(B)
OUT at the
rear.
-
Set the Trigger
Level (SET A/B)
on the
front
to the follo-
wing
values,
and verify the
voltmeter’s
readout:
SET A(B) Readout
Pass/Fail
Input A Input
B
+
5.00
V
+
5 V ± 0.28V
-
5.00
V
-
5
V
±
0.28V
0.00
V
0 V ± 0.03V
Table
2-8
Trigger level outputs
check
Probe
Comp
View
Press the
PRESET key to set the
timer/counter in default
setting.
Figure
2-3
Adjustment of the
counters
probe.
Select
TIME
A-B, non AUTO, XI
attenuation
and
DC
coupling for
both channels.
-
Set the
LF
synthesizer
to 2 kHz square
wave
and 8 Vpp
amplitude.
-
Connect
synthesizer to Input
A via a
BNC-cable.
-
Use an
oscilloscope
to check the signal
at Probe Comp
View, at the
rear.
The square wave will
have
the same
step response
and
4 Vpp amplitude.
-
Select XI 0
attenuation.
-
Check that
the square
wave
is
2 kHz and 0.4
Vpp.
-
Repeat this
test
for channel
B.
Measuring
Functions
Preparation for Check of
Measuring Function:
-
Press the
PRESET
key,
to set the
timer/counter in
the
default setting.
-
Connect a 10
MHz sine
wave signal
with
2.0 Vpp ampli-
tude to input A.
Select the following
settings for the
timer/counten
-
50 £1 input impedance for
A and B
-
Non
AUTO
-
COMA
-
Check
that the
timer/counter
performs
the
correct
meas-
urement, by displaying
the result as
shown
under the
"Display"
column in Table
2-9.
Selected
Function
Action
Display
Pass/
Fail
FREQ
A
10
MHz
FREQC
3)
PER
A
100‘®
s
RATIO A/B
10000000
Select
NEC
SLOPE
B
10000000
RATIO
C/B
00000000
PWIDTHA
50
000"®
s''^
TIME A-B
50 0
00"®
PHASE A-B
180
or
-180'’^
TOT A-B
MAN
0
Deselect
COMA
0
TOT
ST/STOP
counting
TOT
ST/STOP
stop
counting
Select COM A 0
TOT A
JX_TL
B
1
TOT A
^
B
1
Select
POS SLOPE B
0
DUTY FA
0500000
Select AUTO 0500000
RISE/FALL
A
3 0
000"®
S2)
VOLT
MAX/MIN
+100 0
-
1 0 0 0
V
Table
2-9
Measuring
functions
check
1)
Value
depends on the
symmetry
of
the signal.
2)
Exact value
depends on input signal.
3)
if an
C-option is installed.
2-6
Performance
Check,
Measuring
Functions

Related product manuals