EasyManua.ls Logo

Fluke PM6681 - Page 13

Fluke PM6681
153 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference
Oscillators
X-tal
oscillators
are
affected
by a number of external condi-
tions
like
ambient temperature and supply voltage but also
by
ageing.
Therefore it is hard to give limits for the allowed
frequency deviation.
The
user
himself must decide the li-
mits depending
on his application, and recalibrate the oscii-
lator accordingiy. See the Preventive
Maintenance chapter.
Oscillator Max
temperature
dependence
Max
ageing
month
Max
ageing year
Standard.
01
±100 Hz ±5
Hz
±50
Hz
PM
9678B,
02
±10 Hz ±1
Hz
±5
Hz
PM
9690,
04
±0.15 Hz ±0.2
Hz
±1 Hz
PM
9691,
05
±0.05 Hz ±0.1
Hz
±0.75 Hz
Table
2-6
Deviation (for PM 9690 and PM 9691 after
48
hours
warm
up time)
To check the accuracy of the
oscillator
you
must have
a
ca-
librated reference signal that
is
at least
five
times as stable
as the oscillator that you
are testing,
see the following ta-
ble. If you use a
non
10 MHz reference, you can use the
mathematics in
PM
6681 to
multiply the reading.
-
Set the counter to default settings
by pressing
PRESET.
Connect the reference to input A
-
Check the readout against
the
accuracy requirements of
your application.
Acceptance Test
As
an acceptance test
the
following table gives a worst case
figure after 30 minutes warm
up
time.
All deviations that can
occur in a year are added together.
Oscillator Frequency
readout Suitable
reference
Pass
/Fail
Standard, 01 10.00000000 MHz
± 150
Hz
PM
9678B
PM 9678B, 02 10.00000000
MHz
±15
Hz
PM 9690
PM
9690,
04
10.00000000
MHz
±2Hz
PM 6685B
PM
9691,
05 10.00000000 MHz
±1Hz
PM 6685B
Table
2-7
Acceptance test for
oscillators
Resolution Test
-
Connect
a
pulse
generator
to a power splitter.
-
Connect one side
of the
power splitter to the A input of
the
counter
via a coaxial cable.
-
Connect the other side
of the power splitter to the B in-
put of the counter.
Settings for the pulse generator,
-
Amplitude
=
1 Vpp, (high
level
+1V and low
level
OV)
-
Period approximately
1
|lis
-
Duration
=
approximately 50
ns
-
Rise time
2
ns
Settings for the timer/counter, after Preset:
-
Function
=
Time A-B
-
Single
Press STAT key under PROCESS
Press SELECT key until display show ’ST DEV’.
Meas Time
=
50
ills
A and B inputs:
-
50 O input impedance
-
Non AUTO
-
Trigger level
=
0.5V
-
DC
coupling
The result should
be (std
dev)
<
0.05'®
s.
Rear Input/Output
10 MHz OUT
-
Connect an oscilloscope to the 10 MHz output on
the
rear of the counter. Use coaxial cable and 50
Q. termina-
tion.
-
The output voltage is sine wave shaped and
should be
above 500 mV rms
(1
.4 V
p-p).
GATE OPEN Output
-
Set your timer/counter in Default
setting
by
pressing the
PRESET.
-
Select CHECK,
Non AUTO, and
Meas
Time
=
5 ms.
~
Connect the oscilloscope to the Gate Open
output via a
coaxial cable. Set the oscilloscope
to
Ims/division.
-
The Gate Monitor
output should be
a pulse similar to the
Figure
2-2
.
1 Mohm:
50
ohm:
>2.8V >1.4V
<0.4V <0.4V
Ims/div.
Figure
2-2
Signal on
gate open output
REFERENCE IN
-
Set the counter to Default Setting
by
pressing PRESET.
-
Connect EXT REF out from another counter to input A.
-
Connect
a 10
MHz
± 100
Hz, 200 mV
rms,
(0.57
V
p-p)
signal to EXT REF
IN
at the rear, terminated with 50Q.
-
Select Ext Ref.
-
The display should show
10
MHz.
-
Change the input frequency to
5, 2,
and
1
MHz
respec-
tivly.
-
The display should still show 10 MHz.
Performance Check, Rear
Input/Output
2-5

Related product manuals