Q10. How do I know if sufficient blood has been applied to the test strip?
If insufficient blood is applied, an error beep occurs, and an error code appears on
the screen (please refer to the Owner’s manual and check error messages.) Please
review the instructions and repeat the test with a new test strip.
To ensure enough blood has been applied, the front side of the test strip should face
up when inserting the test strip. Test results might be wrong if the contact bar is not
fully inserted into the test slot.
Q11. What does the clinical validation on FORA’s BGMs mean?
The clinical validation is a testament to the accuracy of FORAs blood glucose
monitors. The glucose testing value should comply with the international standard:
EN ISO 15197:2015
Analytical performance evaluation 95% of the measurement to fall:
±15mg/ dL at glucose concentrations100mg/ dL (5.55mmol/L)
±15% at glucose concentrations100mg/ dL (5.55mmol/L)