EasyManua.ls Logo

ForaCare 6 Connect - Troubleshooting; Q1. Troubleshoot abnormal results and error messages

Default Icon
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Back to Top 4
Troubleshooting
Q1. Troubleshoot abnormal results and error messages
Item
Outcome of defect
Root Cause
Solution
1
A significant fluctuation
between measurements.
(1) Using old or expired test
strips.
(2) The blood samples were
obtained from different sites
each time.
(3) Waited too long to insert the
test strip into the device.
(4) The blood samples are
contaminated. (e.g. the
pricked site is not clean)
(1) The strips could be
expired or malfunctioned
due to improper storage.
Please use new strips
within the expiry date.
(2) Please obtain blood
samples from the same
site.
(3) Please insert the test strip
as soon as you collect the
blood sample.
(4) Please wash your hand
before testing.
2
The results might be
higher
(1) Excess plasma in the blood
sample due to excessively
squeezing of the fingertip.
(2) The test strip has expired or
is affected by temperature
and humidity.
(3) Patients health conditions:
a. Medications.
b. Time and dosage of
insulin administered.
c. HCT level.
d. Interference factors in the
diet, such as vitamin C
(ascorbic acid), glutathione
(antioxidant), and uric acid
(for GDH strips).
e. Patient Hydration level.
(1) Conduct the test again.
Do not excessively
squeeze the pricked site.
We recommend wiping off
the first drop of blood and
using the second drop of
blood (to avoid effusion
tissue fluid).
(2) Please open another vial
and use new strips to do
the test again.
(3) Please check with your
doctor to make sure your
status is suitable for
measuring blood glucose
without any interference.
3
Error message: E-U/ Err
24
(1) Insertion of a used test strip.
(2) The blood sample was
inserted before the blood
drop symbol.
Repeat the test with a new
test strip. Wait for the blood
drop symbol to blink, then do
a retest using a proper
amount of blood.
4
Error message:E-b/ Err
36
Battery power is too low for the
meter to operate.
Replace or charge the
batteries before conducting
the test.
5
Error message: E-F
Strip was removed soon after
applying blood and not given
enough time to be absorbed by
the absorption hole
Repeat the test with a new
test strip and wait for the
result.
6
The BGM keeps
beeping until it is out of
the battery.
The Electronic component is
defective in PCBA.
Please contact FORA
customer service for a repair
or replacement.

Table of Contents

Related product manuals