Q27. Where can I find the control solution range?
The range of the control solution test is printed on the test strip vial or individual foil
pack, and your test should fall within this range. If the test result is out of range, read
the instructions again and repeat the control solution test.
Q28. Reasons for abnormal results in control solution tests.
The results might be higher.
Some residue is on the
control solution’s carrier
(cap or tip).
Clean the carrier (cap or
tip of control solution) and
test again. This time use
the second drop of the
control solution.
The results might be lower.
(1) There are pollutants in
the control solution.
(2) Insufficient control
solution is absorbed.
(1) Replace the control
solution and test
again.
(2) Please test again and
remove the strip only
after hearing the
“BEEP” sound from
the meter.
The results might be higher or
lower.
(1) It is not in the correct
user mode (e.g., not in
CTL or QC Mode
during the control
solution test).
(2) You may be using a
different control
solution.
(3) Control solution may
have expired.
(1) Choose the correct
user mode and test
again.
(2) Use ForaCare control
solution only.
(3) Discard after opening
90 days.
Large variance in between tests.
The temperature of the
strips and meter is
unstable.
Please test again after
placing the strips and
meter in the same
environment for 30~60
minutes.