8. FORA ADVANCED proβ-Ketone Test Strip Test Strip
ACS052
Accuracy
Accuracy testing shows that results are comparable between trained
professionals and lay users. Accuracy was assessed with 100 patients by
comparing whole blood β - OHB results with plasma results obtained using
reference laboratory instruments. Please see Table 1.
Table 1 - Accuracy
Mean different plot
0.1 mmol/L
Slope
1.0693
Intercept
-0.0051
R (corr. coef.)
0.9959
Precision
Precision testing shows that results typically vary by 3.06% to 3.33%.
Please see Table 2.
Table 2 - Precision
Low
Mid
Mean mmol/ L
0.487
2.89
SD mmol/ L
0.051
0.088
CV%
-
3.06