9. FORA COMFORT Blood Glucose Test Strip
ACS042
Accuracy
The table below displays how often FORA COMFORT achieves this target. The
chart is based on a study carried out on 160 patients (each patient was tested six
times which had 960 test results) to see how well FORA COMFORT performed
compared to YSI-2300 reference method results.
Table 1 Accuracy results for glucose concentration < 100 mg/dL (5.55 mmol/L)
Within ±5 mg/dL
(Within ±0.28mmol/L)
Within ±10 mg/dL
(Within ±0.55mmol/L)
Within ±15 mg/dL*
(Within ± 0.83 mmol/L)
57.3% (165/288)
90.3% (260/288)
100% (288/288)
Table 2 Accuracy results for glucose concentration100 mg/dL (5.55 mmol/L)
Within ±5 %
Within ±10 %
Within ±15 %*
45.7% (307/672)
77.8% (523/672)
95.4% (641/672)
Table 3 Accuracy results for glucose concentrations between 32.3mg/dL
(1.79mmol/L ) to 544.0mg/dL (30.22mmol/L)
Within ±15 mg/dL or ±15%
(Within ±0.83 mmol/L or ±15%)
96.8% (929/960)
Note: *According to the accuracy criteria of EN ISO 15197: 2015, 95% of all
differences in glucose values (i.e., YSI-2300 reference values minus glucose
values of FORA COMFORT) should be within ±15 mg/dL (0.83 mmol/L) for
glucose concentration < 100 mg/dL (5.55 mmol/L), and within ±15% for glucose
concentration ≥ 100 mg/dL (5.55 mmol/L). When Test Strips results are compared
to the reference values, difference values below 100 mg/dL (5.55 mmol/L) are
expressed in mg/dL or mmol/L, while those above 100 mg/dL (5.55 mmol/L) in
percentage.
User performance
160 subjects were tested with blood samples taken from the fingertip and the
alternative sites, the palm, the forearm and the upper arm. The tables show how

Table of Contents

Related product manuals