EasyManua.ls Logo

FRC TankVision WLA200 - Figure 3. Pressure Sensor

FRC TankVision WLA200
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WLA200 Rev140407
11
Apply sealent around
base and threads.
Thin Wall Adapter
Figure 3. Pressure Sensor
Apply epoxy around
ange, threads, and
screws.
Apply sealent
around base and
threads of sensor.
Typical Pressure
Sensor Location
Typical
Tank
1/2 NPS
2"
Mount sensor approximately
2" from bottom of tank.
Note: The sensor can be mounted
vertically on a 90° tting in cold areas to
help prevent water freezing in the sensor.
0.6"
7/8"
Across Flats
1/4 NPT
2.8"
(max)
1.1"
3-Pin
Packard Plug
0.87"
Note: Do not mount the sensor in the
bottom of the tank. Sediment may collect
in the port and cause sensor failure.

Related product manuals