Page 3
Fresenius 2008
®
K Preventive Maintenance Procedures
P/N 507297 Rev. H
1.2 OPERATING MODES
The following preventive maintenance procedures contain instructions to place the
2008K into Dialysis Mode and Service Mode.
To place the machine in Service Mode, turn the machine power On and wait for the
message Press CONFIRM for Service Mode to appear. Once it appears, press the
[CONFIRM] key and the message will change to Machine in Service Mode. After
the System Initializing process is complete, the machine will be in Service Mode.
If the [CONFIRM] key is not pressed when the Press CONFIRM for Service
Mode message is on the screen, the screen will change and the message Machine in
Dialysis Mode will appear. After the System Initializing process is complete, the
machine will be in Dialysis Mode.
P
rinted 5/18/2012 12:02PM

Other manuals for Fresenius Medical Care 2008K

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008K and is the answer not in the manual?

Fresenius Medical Care 2008K Specifications

General IconGeneral
ManufacturerFresenius Medical Care
Model2008K
TypeHemodialysis Machine
Dialysate Flow Rate300-800 mL/min
Arterial Pressure MonitoringYes
Venous Pressure MonitoringYes
Power Supply50/60 Hz
Ultrafiltration Rate0-4000 mL/h
DisplayLCD
Safety FeaturesBlood leak detection, air detection, pressure monitoring, temperature monitoring, conductivity monitoring
Conductivity Range13.0-16.0 mS/cm

Related product manuals