Fresenius 2008K Preventive Maintenance Procedures
P/N 507297 Rev. A
2.6 ALARM AND PRESSURE HOLDING TESTS
Verify the automatic alarms produce the responses shown in Table 1 and the
machine passes the automatic pressure holding test as follows:
1. Place the dialysate lines in the shunt and close the door.
2. Place a venous chamber filled with water in the holder on the level
detector module.
3. Place the machine in Dialyze mode and start the blood pump. Clear all
blood and water alarms.
4. Select the Test & Options button. On this screen press the Both Tests
button. Press the [CONFIRM] key to start.
5. Observe the machine stepping through the following alarm tests and
ensure that each alarm produces all of the responses shown in Table 1.

Table of Contents

Other manuals for Fresenius Medical Care 2008K

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008K and is the answer not in the manual?

Fresenius Medical Care 2008K Specifications

General IconGeneral
ManufacturerFresenius Medical Care
Model2008K
TypeHemodialysis Machine
Dialysate Flow Rate300-800 mL/min
Arterial Pressure MonitoringYes
Venous Pressure MonitoringYes
Power Supply50/60 Hz
Ultrafiltration Rate0-4000 mL/h
DisplayLCD
Safety FeaturesBlood leak detection, air detection, pressure monitoring, temperature monitoring, conductivity monitoring
Conductivity Range13.0-16.0 mS/cm

Related product manuals