Advanced Features
5-11
Select RE-CALIBRATE. During this 4-minute routine, the
monitor internally corrects deviations from calibration
coefficients set at the factory.
The monitor may determine that the sensor is the cause of the
malfunction. Retry VERIFY ACCURACY after properly
attaching the sensor clip to the verification filter. If this
message continues, it is necessary to contact your service
representative.
Upon successful completion of RE-CALIBRATE, the monitor
will indicate that it is ready for clinical use. It is not necessary
to VERIFY ACCURACY if RE-CALIBRATE is successful.
If RE-CALIBRATE fails, the instrument is unable to re-
calibrate itself to the necessary parameters. If this should
occur, make sure the verification filter is clean and that the
sensor clip has been properly attached (Refer to Section 2).
RE-CALIBRATE may be retried until successful.
If VERIFY ACCURACY and RE-CALIBRATE should both
fail, the CRIT-LINE will prevent you from exiting the
CALIBRATION MENU and will only allow you to select RE-
CALIBRATE. This safety feature prevents patient monitoring
if the monitor is not providing data representative of actual
values.
If the CLM III should fail both the VERIFY ACCURACY and RE-CALIBRATE routines, either the
Sensor clip is improperly attached to the verification filter or the CLM III’s self-diagnostic capabilities
have determined that a problem exists. If you are certain the sensor clip is properly attached to the
verification filter and RE-CALIBRATE continues to fail, contact your service representative.
NOTE: If you wish to verify accuracy after data has been collected (i.e. after a patient run
instead of prior to a patient run), the MEMORY FULL screen will appear. You must erase
all data prior to performing a verify accuracy test. Any data you want to retain must be
printed or downloaded prior to performing the verify accuracy test.
SENSOR MALFUNCTION
RETRY OR REFER TO
THE USER MANUAL
RE-CALIBRATION HAS FAILED
RETRY OR REFER TO THE
USER MANUEL
VERIFICATION FAILED
SELECT RE-CALIBRATE
CALIBRATION DONE
READY FOR
CLINICAL USE
SELECT RETURN
INVALID SELECTION
SYSTEM IS NOT READY
FOR CLINICAL USE!!