MANUAL CONTROL OPERATIONSContinued
28 P/N 450385 Rev. H
9.3 MAIN POWER DISCONNECT
The MAIN POWER Switch is provided to allow the Operator to completely shut down the power to
the GranuFlo I Dissolution Unit. The Main Power Disconnect Switch should be switched to the OFF
position when the GranuFlo I Dissolution Unit is not in use or in case of an emergency.
Remove POWER PLUG from wall receptacle to
disconnect power. A 'LOCKOUT' device may
be used to prevent unauthorized start-up.
NOTE: WHEN POWER IS RESTORED TO THE GRANUFLO I DISSOLUTION UNIT AFTER THE MAIN POWER
SWITCH IS TURNED OFF, OR IF A POWER FAILURE OCCURS, THE GRANUFLO I DISSOLUTION UNIT WILL RESUME
THE OPERATION THAT WAS IN OPERATION AT THE TIME OF THE INTERRUPTION
, AFTER THE OPERATOR HAS
PRESSED THE
START BUTTON.
PRESSING THE PAUSE BUTTON OR PLACING THE CONTROL INTO THE STEP MODE WILL DISABLE THE PUMPS,
AGITATORS
, DRAINS, FILL VALVES, ETC.