TN5A4215b 14/58
[WHITE] When alarm is generated, white backlight flashes.
(5) Sounding time
Sounding time setting is selectable from the following settings for each alarm type.
[Continuous] When buzzer sound setting is ON”, buzzer sounds continuously in alarm condition.
[1 min] to [15 min] When buzzer sound setting is ON, buzzer sounds for set period and then buzzer
stops.
(6) Mute
Mute setting is selectable from the following two settings for each alarm type.
[Available] Buzzer can be stopped by pressing “” button when buzzer sounds.
[Not available] Buzzer cannot be stopped when buzzer sounds.
(7) Vibrator
Vibrator setting is selectable from the following two settings for each alarm type.
[ON] When alarm is generated, vibrator operates.
[OFF] When alarm is generated, vibrator doesn’t operate.
(8) Flash LED
Flash LED setting is selectable from the following two settings for each alarm type.
[ON] When alarm is generated, flash LED on top of the display turns on.
[OFF] When alarm is generated, flash LED doesn’t turn on.
(9) Latch
Latch setting is selectable from the following two settings for “Dose rate alarm/ Dose rate warning”.
[ON] When alarm is generated, alarm operation continues for 10 seconds or continuously according
to the setting if the alarm condition is canceled.
[OFF] When alarm is generated, alarm operation stops if the alarm condition is canceled.