Appendix A: Interface 1000 Specifications--Floating Operation
Appendix A - 3
Potentiostatic Mode
Applied Voltage Range Max ±12 volts
Accuracy
DC zero offset
Gain
Max
1
0.2
mV
% setting
Note 18
DC Bias Typ ±8 volts
Scan DAC ranges Typ ±6.4, ±1.6, ±0.4 volts
Drift Max <20
µV/C
Note 19
Noise and Ripple
1 Hz to 1 kHz
1 Hz to 200 kHz
Typ
10
20
µV rms Note 20
Galvanostatic Mode
Maximum Full-scale Current ±1000 mA
Minimum Full-scale Current 10 nA
Accuracy Dominated by current to voltage
converter accuracy shown above
Sig Gen Voltage for Full-scale Current ±3.0 V
Environmental
Operating Temperature Range 0 to 45
C
Relative Humidity Max 90 (non-condensing) %
Storage and Shipping Temperature –25 to 75
C
Maximum Shipping Acceleration 30 G