Appendix B: Interface 1000 Cell Connector--Floating Operation
Appendix B - 1
Appendix B: Interface 1000 Cell Connector
Chapter 4 describes the connections between a cell cable and an electrochemical cell. This appendix
describes the other end of the cell cable.
Multiple pins assigned to the same signal are connected together on the Interface 1000’s Potentiostat
board. If you need to connect this signal outside the Interface 1000, you need a wire connected to any
one of the D-connector pins.
Table B-1
Cell Connector
Connected to the working electrode in most cases (see Chapter
has a 261 Ω resistor in the cell end of the cable. Custom cell cables are likely
to require a similar resistor.
The shield for the working
Driven to the same potential as Pin 1.
Left open at the cell end of the cell cable.
Shield
The shield for the reference electrode input.
Pin 16. Left open at the cell end of the cell cable.
18, 19, 22,
24
Can be used to shield the cell if very low
currents need to be measured. Also used as a shield for the counter electrode
cable.
With an Interface 1000, used to sense the potential of the counter electrode.
Allows accurate ZRA mode even with significant cell current through counter
lead’s resistance.
Used to identify the type of cell cable attached to
the unit. Pull to a logic High through a resistor. Ground to set the bit low.
ID bits. Used to identify the type of cell cable attached to
the unit. Pull to a logic High through a resistor. Ground to set the bit low.
Shield
The shield for the working electrode inpu
Left open at the cell end of the cell cable.
Electrode
ected to the working electrode.
The cell current flows through this pin.
Electrode
Connected to the reference electrode in most cases (see Chapter
lead has a 261 Ω resistor in the cell end of the cable. Custom cell cables are
likely to require a similar resistor.
ID bits. Used to identify the type of cell cable attached to
the unit. Pull to a logic High through a resistor. Ground to set the bit low.
ID bits. Used to identify the type of cell cable attached to
the unit. Pull to a logic High through a resistor. Ground to set the bit low.
Electrode
s counter electrode. The cell current flows through this
connection.
Designing cell cables for a 1 MHz potentiostat is not a trivial task. Gamry Instruments does not
recommend user-designed cables (except as a last resort). In many cases we can build a custom cable to
meet your specific needs. We’ll also be happy to discuss your cable requirements and give you assistance if