Appendix A -- Reference 3000 Specifications--
11 - 1
Appendix A -- Reference 3000 Specifications
All specifications are at an ambient temperature of 25 °C, with the Reference 3000 powered using the power
adapter shipped with the unit, standard shielded 60 cm cell cables, and the cell enclosed in a Faraday shield.
All specifications are after software calibration.
A numbered note qualifies many of the specifications. Many of these notes describe the method used to
measure a specification. The notes are found at the end of this appendix.
All specifications are subject to change without notice.
Control Amplifier
High Voltage, Low Current Mode
Compliance Voltage Min
Typ
± 30
± 31.5
volts Note 1a
Output Current Min
± 1500
mA Note 2
Low Voltage, High Current Mode
Compliance Voltage Min
Typ
± 15
± 16.2
volts Note 1b
Output Current Min
± 3000
mA Note 2
Both Modes
Unity Gain Bandwidth Typ 1100, 330, 50, 5.0, 0.5 kHz Notes 3, 4
Slew Rate Typ 10, 3, 0.5, 0.06, 0.006
V/µsec
Notes 3,4
Differential Electrometer
Max Input Voltage Min
± 10
volts Note 5
Input Current Max 6 pA Notes 6
Input Resistance
Differential (between inputs)
Common Mode (input to ground)
Typ
100
1
T
Note 7
Input Capacitance
Differential (between inputs)
Common Mode (input to ground)
Typ
0.2
12
pf
Note 7
Bandwidth (-3 dB) Min 15 MHz Note 8
CMMR
DC to 100 kHz
100 kHz to 1 MHz
Min
80
60
Note 9
Voltage Measurement
A/D Full Scale Ranges Typ
±12.0, ± 3.0, ± 0.3, ± 0.03
Volts
Note 3
Resolution Typ 400, 100, 10, 1
µV/bit
Zero Offset Error Max 1 mV Note 11
Gain Error Max 0.2 % Note 11