EasyManuals Logo

Gantner Q.bloxx A108 User Manual

Default Icon
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
Vers. No. 6.1
112 Released: 25/04/2017
5 Configuration Configuring analog inputs
Fig. 5-4 Setting up zero balance or tare.
When the variable (Zero_tare in this example) has certain values,
taring or zero balancing is carried out or an existing zero or tare
value is deleted (undo zero balancing/taring):
Do not forget to save these settings in the module and the Test
Controller (refer to Section 5.3.1, Setting sensor parameters,
pag
e 107).
5.3.4 Sensor scaling over several points (table)
In the standard setting (linear) sensors are scaled via a charac-
teristic curve with two points. You can however also perform scal-
ing over several points (table). During the measurement the val-
ues are then linearly interpolated between the specified points.
Procedure 1. Create the sensor as usual (two-point scaling).
2. In the Sensor c
olumn click on the sensor.
3. Copy the sensor (click on Copy) and sp
ecify a name, e.g.
ForceTable (Fig. 5-5). If the scaling to be entered later only
applies to this sensor, you should use the sensor type and the
serial nu
mber as the name, e.g. C2 #1234 (5V).
Value Function
1 Perform taring when setting the value
2 Reset tare (delete tare value)
4
Start zero balance The mean is formed until the value 0
is written to the variable again.
8 Reset zero balance (delete zero value)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gantner Q.bloxx A108 and is the answer not in the manual?

Gantner Q.bloxx A108 Specifications

General IconGeneral
BrandGantner
ModelQ.bloxx A108
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals