Installation
3-17
3. Workstation Inspection
Inspect each workstation component listed under the heading workstation inspection. Select one check
box for each item in the list.
WorkStation Inspection list:
found
parts
damaged parts
scratched paint
3.4.3. Internal Inspection
Observe electrostatic discharge safety procedures.
Removing any covers exposes dangerous electrical voltages on several components.
When possible, disconnect system from AC line before removing any covers. When it is
necessary to leave the power applied, use extreme care to avoid death or injury from
electrical shock.
1. Remove workstation cover. Inspect the internal labels (refer to chapter 1.2.11) and parts. Check if
there are loose parts.
2. Remove the C-Arm’s covers. Check all cables and PCBs inside C-Arm. Look for loose PCBs and
connectors and reset.