EasyManua.ls Logo

GE Brivo OEC 865 - Cleaning and Disinfections

GE Brivo OEC 865
551 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation
3-19
The system can adapt power supply of AC100V, AC110V, AC115V, AC120V, AC200V,
AC220V, AC230V and AC240V by different connections of the transformer. Before install
the system, please check the power supply voltage range and the transformer
connection per the diagram below.
Wire Transformer taps
Voltage
240 V
230 V
220 V
200V
120V
110V/115V
100V
Wire N (Brown)
1
1
1
1
1
1
1
Wire L1 (Blue)
5
4
4
3
5
4
3
Wire N1 (Black)
2
2
2
2
2
2
2
Wire L (Red)
8
8
7
6
8
7
6
Jumper
10-11
10-11
10-11
10-11
9-10
11-12
9-10
11-12
9-10
11-12
3.5.2. UPS Setup
UPS is assembled in back of workstation. Considering safety of shipping, battery is not connected with
power supply.
Follow these steps to setup the UPS.
1. Open the Power Distribution package.
2. Release screws to fixe UPS. Then take out UPS.