Loading...
Periodic Maintenance
8-13
8.6.3. Workstation Checks
Mechanical Movement:
Refer to 3.6.1 for Workstation mechanical movement check.
Adjust the monitor vertical movement annually, according to below procedure:
10N<Force of monitor vertically movement<50N
Monitor vertical movement PM:
1. For 865 need remove monitor back cover.
2. Screw the nut to adjust the tightness of the monitor movement in moderate strength.
Brivo OEC 715/785 Back View
Brivo OEC 865 Back View
If friction of monitor horizontal movement is not effective, perform below procedure to
refresh the friction of monitor horizontal movement.
10N<Force of monitor horizontal movement<50N
Monitor horizontal movement PM:
1. Remove screws on two sides of monitor support
cover.
2. Remove 2 inside screws.
3. Rotate the secure nut clock wisely to increase
friction for horizontal rotation