Calibration
4-51
Camera Focus Calibration
Camera Focus adjustment is done in the factory. Adjust with much care only if
necessary.
1. To have access to CCD Camera Focus, perform the following.
a. Turn off the power.
b. Position the C-arm to lateral position.
c. Disconnect the Camera Cables.
d. Remove counter weight.
e. Reconnect the Camera Cables.
f. Power on the system.
2. Loosen Camera Focus Securing Screw, adjustment Camera focus.
3. Make an exposure, and check the resolution again. Repeat as many times as necessary.
4. If passed:
a. Tighten the Camera Focus Securing Screw.
b. Power off system
c. Disconnect camera cables
d. Install counter weight
e. Connect camera cable
f. Install I.I. cover
5. If the resolution check still fails after many trials, proceed to I.I. Focus adjustment.
Camera of Brivo OEC 715/785 5075006 Component GigE Camera AM1565
Camera Focus
Adjustment
Ring
Camera Focus
secure Screw