Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
 
5-12 
 
System provide tool to set imaging parameters. Click “Image Processing configure” tab on the main 
screen, the image processing interface is displayed. 
Changes on Image Brightness, Image Contrast, Black/Wihte Margin and Sharpen 
power are not recommended because these values have been optimized for the image 
quality to compromise any anatolomical imaging. Any changes will affect the system 
image quality. 
•  B/C Preference  High Brightness or High Contrast: According to user’s  preference  to select 
brightness enhance method. 
•  Image Brightness: Pressing up or down arrow or dragging the block allows you to adjust image 
brightness default from 0 to 100. 
•  Image Contrast: Pressing up or down arrow or dragging the block allows you to adjust image 
contrast default from 0 to 100. 
•  Edge Enhance: Select default value of sharpening power from 0-9. 
•  Rotation Angle: Select default value of image rotation angle from 0~359. 
•  TNR (Temporal Noise Reduction): Select default value of noise filter from 0-3. 
•  SM (Smart Metal) value: Select default value of the appearance reduction of metal in the image 
from 0-50. 
•  Enable or disable IP circular mask: it is used to assist beam alignment calibration. Make sure it is 
set back to the default setting “Enable” after beam alignment. 
•  Enable or disable IP shading correction image processing function.